Dr N.I. Leaves wrote:
>> To follow up: my laser has been replaced. But now Applied Biosystems
>>is saying that the "block dropping" phenomenon probably has nothing to
>>do with the laser. Rather version 2 of data collection can start to drop
>>blocks in the way I describe above if the sequencer and computer are not
>>restarted every 2-3 days!
>> Given that version 2 is set-up to allow continuous running (rolling
>>loading) this seems most unfortunate. Restarting the sequencer wastes
>>all the polymer that is in the sheath flow shringe.
>> Has anyone seen this phenomenon?
>>Phillip SanMiguel
>>Purdue Genomics Core Facility
>>>>> Hi Phillip,
>> our three 3700 machines are run pretty much constantly, I reckon we
> restart the computer and sequencer about once every two weeks. I have
> never seen the dropping out you describe.
>> nick
>>[...]
Thanks for the response, Nick.
Are you running version 2.0 of Data Collection?
Applied Biosystems is now saying that restarting once a week should be
sufficient.
Phillip SanMiguel
---