>>>>> "Marc" == Marc Woodland <m.p.woodland at ic.ac.uk> writes:
Marc> You've installed the new software all seems to be O.K.
Marc> your gel is loaded you go to the start icon and and error
Marc> message comes up "Flow detected with pump off"
Marc> DO NOT PANIC
Marc> Go to "Window" and open up Mannual control. Select "Pump
Marc> On" and for value insert 51, for range 0-60 then execute.
Marc> Then select "Pump Off". repeat this cycle a few times then
Marc> try to restart your run. You may need to do this more than
Marc> once. I had three gels run under these conditions the last
Marc> time I think I mannually 'switched the pump on/off about 15x
Marc> before it ran.
Marc> If this doesn't work you could reinstall the old software or
Marc> BURST INTO TEARS and ring your engineer for some
Marc> counselling.
This is a check that was added with XL firmware version 2.0, and it
indicates a faulty polymer flow detector in the instrument. The
detector should thus be changed by your field service representative.
We had a debate over whether to allow the user to ignore this condition
and start the run anyway in version 2.1 (because so many people had
received the above message and gotten confused), but the overriding
reason for not doing so was to ensure that faulty equipment gets
replaced. (This was partly for marketing and/or legal reasons as
well).
Cheers,
Tor Slettnes,
Firmware Engineer,
PE Applied Biosystems