hello,
it´s not good to dry your column. to take the air off you need to increase
the pressure inside the column, so close the exit for a few seconds,
respect the maximum pressure of the column.
after this, you need to check your column, to see if it working properly.
best regards,
Joana Arroz
2016-01-10 23:32 GMT+00:00 Bilal Majed <majedbilal from gmail.com>:
> Hello
> After I had completed my analysis, I switched to sequence flush overnight
> for two hours. However I made an error in the settings and realised in the
> next morning the method was still running and I ran out of wash phase. So
> air was being pumped into the system and the column.
>> Please can you help me correct this mistake.
>> Hplc system- Agilent 1220 Infinity LC System
> Column- Inertsil ODS 3V
> Reverse phase system
>> Thank you
>> Bilal Majed
>> Sent from my iPhone